Lisa M
A compact, fully integrated ELISA workstation tailored for low- to medium-throughput laboratories, designed to enhance assay accuracy, reproducibility, and consistency in immunodiagnostic testing.
• Compact benchtop model for ELISA workflow management
• Fully integrated system for automatic processing of sample-loaded microplates
• Four dedicated dispensing channels with independent precision pumps for reagent dispensing
• Reduces assay drift for improved reliability and reproducibility with high speed of reagent dispensing
• Integrated orbital and reciprocal shaker delivers effective plate shaking within 50–1200 RPM range
• Inbuilt incubator for Consistent and evenly distributed heating
• High-efficiency washer module with versatile operating modes to suit diverse assay requirements
• High-performance LED Photometer with a dynamic range of 0–4.0 OD for accurate, high-sensitivity readings
• Intuitive interface designed to make every operation effortless and smooth
• No consumables required – no hidden costs, ensuring cost-effective operation over time
Technical Specifications
General Specifications
-
Sample Format: 1 × 96-well microplate (ANSI SLAS compliant)
-
Dimensions (H × W × D): 560 × 455 × 390 mm
-
Weight: 25 kg
-
Printer: Inbuilt thermal printer and external EPSON LX300 (ESC/POS protocol supported)
-
LIS Connectivity: Unidirectional proprietary protocol*
-
Power Requirement: 150 W
-
Operating Temperature: 10–35 °C (50–95 °F)
-
Operating Humidity: 40–80% (non-condensing)
Incubator
-
Temperature Range: Ambient +4 °C to 55 °C (1 °C increments)
-
Temperature Stability: ±1 °C across the plate at 37 °C
-
Incubation Time: Programmable, 1–999 minutes
-
Temperature Monitoring: Real-time monitoring of room temperature and incubator temperature
ELISA Reader
-
Dynamic Range: 0–4.0 OD at 450 nm
-
Light Source: LED
-
Spectral Range: 400–750 nm
-
Filter Slots: 10
-
Pre-installed Filters: 405, 450, 492, 620 nm
-
Photometric System: Single photometric channel
-
Precision: ≤1% (0.0–3.5 OD)
-
Accuracy (450 nm): ≤1% (0.0–3.5 OD)
-
Linearity (450 nm): Up to 3.5 OD
-
Drift (450 nm, 12 hours): < 0.001 OD
-
Resolution: 0.001 and 0.0001 OD (user-selectable)
Dispenser
-
Type: Four independent precision pumps
-
Dispense Volume Range: 20–200 µL
-
Precision: <3% at 100 µL
-
Accuracy: <5% at 100 µL
-
Pipetting Time: 96-well plate in ~2 minutes at 100 µL
-
Reagent Compatibility: Universal – compatible with any bottle or vial type
Washer
-
Manifold Configuration: Autoclavable 8-way manifold with two parallel needles
-
Dispense Volume: 100–500 µL (1 µL increments)
-
Wash Modes: Standard, Overflow, Soak, Sweep
-
Residual Volume: <5 µL
-
Soak Time: 0–250 seconds
-
Wash Buffer Capacity: 2 L
-
Waste Buffer Capacity: 2 L
Shaker